DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
267
Figure 11-29 The Waveform Averaging screen
It is important to note that the effective measurement rate when using the averaging
software is not equal to the PRF as it is in standard mode. The effective measurement
rate when using the averaging option is equal to the total PRF divided by the
averaging factor.
At higher averaging rates, the screen update rate may be less than 60 Hz (depending
on the PRF value). If the screen update falls below 60 Hz (the industry standard), an
icon (
) displays in the flag area of the screen (see Figure 11-30 on page 268).
xAv flag on screen