Olympus ENDO-AID Quick Reference Manual Download Page 2

Computer-Aided Detection (CADe) Application

Getting to Know ENDO-AID

ENDO-AID CADe is a high-performing computer-aided detection application for colonoscopy that uses artificial 
intelligence (AI) to suggest the potential presence of lesions such as colonic polyps, malignant neoplasms and 
adenomas.

Switching the System On

Turn the system on by 

pressing the power switch. 

The power indicator lights 

up. Keep the system on for 

all procedures even if the 

CADe functionality will not 

be used.

CAD Indicator Light

Shows that CADe is active

Mode 2
Mode 1

Menu

Directional 

Arrows

Back

91990

Activating the CADe Application 

Press the CAD button or the Mode 1 button 
to turn on the CADe application. To turn it off, 
either push the CAD button once or press the 
Mode 1 button for few seconds.

Portable 
Memory Port

Enter

Summary of Contents for ENDO-AID

Page 1: ...Quick Reference Guide 91889 Endoscopy CAD System OIP 1 ...

Page 2: ...hing the System On Turn the system on by pressing the power switch The power indicator lights up Keep the system on for all procedures even if the CADe functionality will not be used CAD Indicator Light Shows that CADe is active Mode 2 Mode 1 Menu Directional Arrows Back 91990 Activating the CADe Application Press the CAD button or the Mode 1 button to turn on the CADe application To turn it off e...

Page 3: ...hen scope is connected 2 and save the changes 3 Do the same for the option CAD auto OFF so that the system turns off when the scope is disconnected The CAD icon is visible when the CADe application is active A CAD icon with a cross indicates that the application is in standby for example when NBI or RDI is active As soon as the user goes back to WLI or TXI the CADe application will switch on autom...

Page 4: ...c image Findings are surrounded by a marker in a subimage The application is thought to be operated preferentially in Normal Mode If the users want a better visualization of a certain marker they will shortly toggle to Target Mode per default the system will then return back to Normal Mode a few seconds afterwards Users willing to use Target Mode preferentially should select unlimited as the Targe...

Page 5: ...onfigured according to user preferences Configurations for Normal Mode Length and thickness of the flags 1 2 Subimage position and size 3 4 Marker thickness in subimage 5 Maximum number of markers shown up to 10 in the subimage 6 Configurations for Target Mode Marker thickness 7 Maximum number up to 10 of markers shown 6 Duration of Target Mode after transitioning to it 8 e g setting it to unlimit...

Page 6: ...witch section of your CV 1500 user presets you can set the scope switch of your preference to control your ENDO AID CADe Look for the following options Option dev control 1 Emulates the CAD button i e to turn CADe on and off Option dev control 2 Emulates the MODE 1 button i e to toggle between visualization modes It is suggested to use Option dev control 2 on switch 2 This way the user can easily ...

Page 7: ...cation will restart in Normal Mode so that the user must additionally push the Mode 1 button in order to activate Target Mode again To avoid this additional last step a work around would be to toggle between Normal Mode and Target Mode instead of switching the application off completely In this case it is recommended to configure the visual settings of Normal Mode in a way that the subimage is swi...

Page 8: ...s who wish to see more polyps even very small ones use Type A Users who want a lower sensitivity select Type B The preferred detection type can be configured in the Detection type section of the Marker flag menu of the settings Detection Type Description Type A Detects more potential colorectal polyps than Type B Type B Tends to suppress more false detection than Type A ...

Page 9: ...1500DL I CF HQ1100DL I CF XZ1200L I EVIS EXERA III Video Colonoscope CF HQ190L I CF H190L I CF H185L I PCF PH190L I PCF H190L I PCF H190DL I PCF H190TL I PCF HQ190L I EVIS LUCERA ELITE Video Colonoscope CF HQ290ZL I CF HQ290L I CF H290L I CF H290ECI PCF H290ZL I PCF H290DL I PCF H290TL I PCF H290L I ...

Page 10: ...cedure steps and or product information It is not meant to replace the instructions for use Any user of this product must at all times observe all mandatory information for the product found in particular on the labels and the instructions for use ENDO AID CADe is a supporting tool and the user should not rely solely on the device for detection of abnormalities LINK OUT terminal OEV321UH SDI OUT1 ...

Reviews: