26
Problem
Cause
Remedy
Page
Still images cannot be recorded.
The DP2-TWAIN or cellSens /
OLYMPUS Stream is processing
recording.
Wait until the recording processing
completes before starting recording
of the next image. In certain cases, it
may be required to press the Cancel
button on the status bar and record
the image again.
–
The cellSens / OLYMPUS Stream is
processing file save, etc.
Wait until processing completes before
starting recording of the next image.
–
The computer memory is insufficient. Exit from other software before retrying
recording.
Save the images which you did not
save.
–
Picture is too bright.
Exposure correction is set in the +
direction.
Set the desired exposure correction
value.
–
The metering area is set to a dark
area outside the region of interest.
Move the metering area to the area
where you want to obtain optimum
exposure.
–
AE lock, which was set when the
exposure time was longer than the
currently required exposure time, is
active.
Cancel AE lock.
–
The input highlight level adjustment is
too low.
Reset the current level adjustment and
adjust the optimum level again.
–
The microscope illumination is too
bright.
Reduce the microscope illumination
intensity or engage an ND filter to
reduce brightness.
–
DP26