background image

103076-01EN [U8998318], Rev. C, June 2014

Safety Information    9

 

— Enclosing the beam working area with protective panels (for example, 3.0 mm stainless steel can attenuate

the beam to background levels)

Contact your Olympus representative for assistance and suggestions on interlocks and applications that limit 
radiation exposure.

Trigger issues
“Deadman trigger” mode requires the user to pull and hold down the trigger for the duration of the test. 
Releasing the trigger immediately aborts the test.

3.3

Indicators and Statuses

This section explains the DELTA handheld XRF analyzer’s indicators (see Figure 3-1 on page 9) and statuses.

Figure 3-1  Power switch and X-ray indicators

3.3.1

Power Switch with Integral Indicator Light

The power switch is located at the upper rear of the analyzer (see Figure 3-1 on page 9).

POWER ON

Press the I/O switch to turn on the power. 
A green LED indicator illuminates. 

This switch DOES NOT turn the X-ray tube on. No tube power is supplied until the Olympus software is launched.

POWER OFF

Press and hold switch for more than three seconds.
The instrument powers off (see section “Exit Options” on page 17 for more exit options).

3.3.2

X-Ray Indicator

The X-ray indicator is located on the upper rear of the instrument (see Figure 3-1 on page 9). It consists of a six-element red 
LED array and provides two key functions:

X-ray indicator continuously ON (solid red LED array)

This signifies that:

The  X-ray  tube  is  enabled.

There is no radiation exposure to you or bystanders. 

The instrument can be carried or set down safely in this condition.

X-ray indicator flashing ON (blinking red LED array)

This signifies that:

The X-ray tube is powered to full operational level.

The analyzer is emitting X-ray radiation through the measurement window.

In this condition, the analyzer must be pointed towards a test sample.

Power switch

X-ray indicator

Summary of Contents for DELTA 50

Page 1: ...us product safely and effectively Before using this product thoroughly review this instruction manual Use the product as instructed Keep this instruction manual in a safe accessible location Models DE...

Page 2: ...e instrument 1 Locate and remove the shipping papers and documentation 2 Open the carrying case and then remove the DELTA and all of the components Please note that there are two layers of protective...

Page 3: ...omponent Key DELTA All models except DELTA Element and DELTA Inspector Foam top layer 1 Carry case 2 DELTA analyzer 3 Docking station charger 4 USB cable 1 see page 6 for details 5 USB cable 2 see pag...

Page 4: ...del shown 2 Probe 3 Measurement window Prolene or Kapton film 4 Hinged window plate 5 Docking station connector 6 Trigger 7 Handle Non slip rubber grip 8 Battery boot 9 Data port with rubber cover 10...

Page 5: ...DELTA Element and DELTA Inspector models This battery charger is described in section 4 2 2 on page 11 Component Key DELTA All models except DELTA Element and DELTA inspector Docking station empty 1 A...

Page 6: ...omponent key DELTA All models standard on the DELTA Element and DELTA Inspector Battery charger empty 1 Input power 12 VDC 2 Battery cradle and connectors 3 Charging error indicator 4 Battery charging...

Page 7: ...and either Secure it with plastic zip ties or Pack the transport case within another box USB cable 2 Mini USB B to USB A connectors AC power adaptor Provides DC power to docking station Input 110 240...

Page 8: ...accidental exposure the filter wheel goes to position 0 and the X rays shut off The tube current is reduced to 0 0 A and the red light stops blinking Also if the probe nose is pulled away from the sam...

Page 9: ...e I O switch to turn on the power A green LED indicator illuminates This switch DOES NOT turn the X ray tube on No tube power is supplied until the Olympus software is launched POWER OFF Press and hol...

Page 10: ...percent of charge remaining ranging from less than 25 to 100 If a battery has a charge of less than 25 use the docking station to perform a full charge Figure 4 1 Battery charge status button 4 2 Char...

Page 11: ...y A battery hot swap capability is a standard feature with the DELTA handheld XRF analyzer You can remove and replace a battery without having to shut down restart or perform a Cal Check A shutdown st...

Page 12: ...ld the analyzer upside down by the handle with the bottom of the analyzer base pointing upward and the nose pointing away from the operator 2 Pull the rubber latch and lift the cover 3 Pull the tab to...

Page 13: ...up radiation safety and initialization screens see chapter Operation on page 15 The main operations are accessible from the Home screen see Figure 5 1 on page 13 Refer to the User Interface Guide P N...

Page 14: ...y Spectrum Display Chemistry Exit screen and save changes Exit screen and don t save changes Display Expand Display Minimize X Review S W Title Name of screen and current mode Padlock trigger indicato...

Page 15: ...safety notice see and then confirm that you are a certified user System initialization begins immediately The analyzer launches a test screen using the last selected mode see Figure 6 1 on page 15 4 I...

Page 16: ...6 3 on page 16 Figure 6 3 Results 11 At the end of the session export the results to a PC using the data port and USB cable 12 When testing and exporting are complete turn off the analyzer using the...

Page 17: ...ter confirmation 3 Relaunch Restart the application All user interface UI screens have a time out power saving feature that causes the screen to go blank if the UI is not accessed or if the instrument...

Page 18: ...103076 01EN U8998318 Rev C June 2014 18 Operation...

Page 19: ...2014 DELTA Radiation Profile 19 7 DELTA Radiation Profile This chapter includes the radiation profiles for the DELTA handheld XRF analyzers Figure 7 1 DELTA radiation profile Handheld instruments 61 c...

Page 20: ...4 Beam RoHS 2 alloy PVC Blank 50e e All 50 kV readings are taken with standard probe shield in place 80 Copper 10 840 400 250 15 EC 680K 16 1800 800 500 30 71X SR2 solder BK 5 1 BK BK Alloy Mining 31...

Page 21: ...200 120 60 22 7 EC 680K 60 9 1500 1000 1500 1000 750 200 56 37 71X SR2 solder BK BK BK BK BK BK BK BK BK BK Al 319 AA 1 BK 25 15 10 10 6 5 BK BK Alloy Mining 316 stainless 35 100 Aluminum BK BK BK BK...

Page 22: ......

Page 23: ...oltage warning symbol This symbol is used to alert the user to potential electric shock hazards greater than 1000 volts All safety messages that follow this symbol shall be obeyed to avoid possible ha...

Page 24: ...otify the carrier making the delivery of any damage because the carrier is normally liable for damage during shipment Retain packing materials waybills and other shipping documentation needed in order...

Page 25: ...103076 01EN U8998318 Rev C June 2014 Important Information Please Read Before Use 27...

Page 26: ...MA 02453 USA www olympus ims com Printed in the United States of America Copyright 2011 2013 2014 by Olympus All rights reserved All brands are trademarks or registered trademarks of their respective...

Reviews: