PRODUCT SPECIFICATIONS
3. Using
Conditions
Use in medical facilities under the direction of doctor.
Don't use directly to the heart. (BF instrument)
Ground the outer casing for safety.
Don't use in an inflammable atmosphere.
Don't use with the following electronic devices.
a. Among the devices designed to apply medical treatment with electronics to a patient, those
which are not confirmed in safety when combined with the CLH-SC.
b. Among the devices not designed to apply medical treatment with electronics to a patient,
those which are not confirmed in safety such as a leakage current, etc.
Power supply
Rated voltage
Input power supply
Frequency
Voltage fluctuation
100 V
1.7 A
120V
1.7 A
220, 230, 240 V
1.7 A
50/60 Hz
± 10% or less
Environmental conditions
Ambient temperature:
Relative humidity:
Atmospheric pressure:
1-3
10 - 40°C
30 - 80%
700 - 1060 hPa