DISASSEMBLING/ASSEMBLING PROCEDURE
5. DISASSEMBLING/ASSEMBLING PROCEDURE
1. Precautions
on
Disassembling/Assembling
Replace the parts and wires to the original positions.
For electrical safety and standard, be sure to reassemble the following parts to the original
states.
Insulation tube and mylar sheet
Cables clamped and separated from the heating parts or high-voltage parts
Cover screws with a toothed lock washer to suppress a radiation noise
Use the specified parts.
The parts used in this unit are designed protective against vibration, heat and high voltage. Be
sure to select the parts with the same characteristics from the parts list when replacing the
parts.
Be careful when disconnecting the cable housing.
Don't pull the cable. Be sure to use the special tool.
2. Jigs
and
Tools
No.
1
2
3
4
Name
Phillips screwdriver No. 2
No.
OT0376
Specification
Remarks
3. Assembling
Reverse the disassembling procedure.
5-1
Summary of Contents for CLH-SC
Page 1: ...MAINTENANCE MANUAL CLH SC...
Page 10: ...2 2 DESCRIPTION OF BLOCK UNIT NAME BLOCK DIAGRAM MODEL CLH SC...
Page 11: ...3 CIRCUIT DIAGRAM CIRCUIT DIAGRAM 3 1 UNIT NAME 120V DATE 99 01 MODEL CLH SC PAGE 1 4 120V...
Page 12: ...3 2 220V CIRCUIT DIAGRAM UNIT NAME 220V DATE 99 01 MODEL CLH SC PAGE 2 4...
Page 13: ...CIRCUIT DIAGRAM 3 3 230V UNIT NAME 230V DATE 99 01 MODEL CLH SC PAGE 3 4...
Page 14: ...CIRCUIT DIAGRAM 3 4 240V UNIT NAME 240V DATE 99 01 MODEL CLH SC PAGE 4 4...
Page 26: ...6 1 COVER SHASSY CLH SC FIG 1 3 6 EXPLODED PARTS DIAGRAM EXPLODED PARTS DIAGRAM...
Page 27: ...6 2 CLH SC MAIN UNIT FIG 2 3 EXPLODED PARTS DIAGRAM...
Page 28: ...6 3 CLH SC CABLE FIG 3 3 EXPLODED PARTS DIAGRAM...