Safety Check Instruction

 

 

 

 

 

 

 

©  Celon AG 

Page 3 / 9 

 

Issue date: 01.09.2008 

SCI.991007-1.4 

Safety Check 

Instruction.doc 

No. 

Procedure 

020 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

       

 

 

030 

Unit and accessories are externally undamaged: 
- Enclosure and front panel have no serious destructions 

040 

Lift up the unit and shake it gentle. No clatter noise should be audible. 

 

Documentation

 

050 

Docu

ment the test results in “SCR.991007. 

 

Sample of symbol:

  Connection for equipotential bonding 

 

Visible on the back of the unit (see fig. 1) 

Samples of label for power fuses: 

Visible on the back of the unit (see fig. 1) 

 

Sample of symbol: 

“Non-ionizing radiation” 

 

Visible on the back of the unit (see fig. 1) 

Sample of label for volume switch: 

Visible on the back of the unit (see fig. 1) 

Sample of label:

 

Product name 

 

Visible on the front panel (see fig. 2) 

 SPEAKER VOLUME 

 

HIGH 

 

LOW 

Sample of symbol: 

“Waste electrical and electronic equipment” 
Visible on the back of the unit, 

only applicable for 

units

 

within the European Union and sold after 

August 13th, 2005.

 (see fig. 1) 

 

Summary of Contents for CelonLab ENT

Page 1: ...maintenance take care of the different hardware versions which may be applicable see below The hardware version of the unit can be identified by the serial number which can be found on the type plate...

Page 2: ...ountry medical device information Medizinproduktebuch 020 Labels present and legible Sample of symbol Unit is defibrillator safe Visible on the front panel see fig 2 Sample of symbol Caution Read the...

Page 3: ...nnection for equipotential bonding Visible on the back of the unit see fig 1 Samples of label for power fuses Visible on the back of the unit see fig 1 Sample of symbol Non ionizing radiation Visible...

Page 4: ...Celon AG Page 4 9 Issue date 01 09 2008 SCI 991007 1 4 Safety Check Instruction doc Fig 1 Sample of the backside of the CelonLab ENT power control unit Fig 2 Sample of the front panel of the CelonLab...

Page 5: ...Documentation 110 Document the test results in SCR 991007 3 Power output No Procedure Test equipment Electrosurgical analyzer Example QA ES Metron Connection cables HF output banana to the electrosurg...

Page 6: ...l analyzer according to the analyzer s instructions for use Set the power level of the unit to 12 and measure the following parameter Attention Start check procedure if not already done after setting...

Page 7: ...loaded and unloaded at rated load Connect the unit via the power cord with a rated supply voltage Connect the footswitch with the unit 280 Connect the load resistor 100 with the HF output 1 and 2 and...

Page 8: ...ety tester Example Unimet 1000 ST Bender Load resistor 100 25 W or more 5 tolerance low inductive part short time load Connection cables HF output banana to the load resistor Test accessories Power co...

Page 9: ...c 8 Inspection label No Procedure 430 Cover or exchange an inspection label as shown below at the enclosure back of the unit and mark the due date of the next periodic safety check month year The unit...

Reviews: