
8
CELON ELITE QUICK REFERENCE GUIDE
C E L O N P R O C E D U R E
MODE OVERVIEW
Technique
Effect
Mode
Max
Power
VP
Short Explanation
Bipolar
Coag
Fine RFITT
40 W
250
Automatic end-of-procedure detection.
Audible feedback.
Pure RFITT
40 W
150
Automatic end-of-procedure detection.
Audible feedback.
Strong RFITT
(w. and w/o
RCAP)
40 W
260
Automatic deactivation when the tissue
resistance exceeds a limit value.
(Resistance-Controlled Automatic Power (RCAP)
to avoid premature tissue desiccation.)
Soft Coag
120 W
260
Little carbonization and adhesion.
Autostart selectable.
Cut
Fine Cut
120 W
470
Low spark intensity for small thermal effect.
Pure Cut
120 W
530
Medium spark intensity
for medium thermal effect.
Strong Cut
120 W
560
High spark intensity for large thermal effect.
Monopolar
Coag
Soft Coag
120 W
250
Little carbonization and adhesion.
Forced Coag
120 W
2000
Spark allows coagulation
also with relatively small electrodes.
Cut
Fine Cut
120 W
580
Low spark intensity for small thermal effect.
Pure Cut
120 W
630
Medium spark intensity
for medium thermal effect.
Strong Cut
120 W
680
High spark intensity for large thermal effect.
Pulse Cut
120 W
690
Intermittent cutting,
e.g. for endoscopic operations.