BX46
Problem
Lamp
Cause
Remedy
Page
Field of view of one eye does not
match that of the other.
Common
Interpupillary distance is incorrect.
Adjust interpupillary distance.
11
Common
Diopter adjustment is incorrect.
Adjust diopter.
19
Common
Different eyepieces are used on
left and right.
Change one eyepiece to match
the other so that both sides are
the same type.
33
Common
Your view is not accustomed to
microscope observation.
Upon looking into eyepieces,
try looking at overall field before
concentrating on specimen range.
You may also find it helpful to
look up and into distance for a
moment before looking back into
microscope.
–
Image shifts when you touch
stage.
Common
Stage is not properly mounted.
Clamp stage.
18
X- and Y-axis knobs are too tight
or too loose.
Common
Tension of X- and Y-axis knobs is
too high or too low.
Adjust tension.
18
Stroke has reduced.
Common
Stage guide is deviated.
Correct deviation as described.
18