Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-BI30 Binocular Viewing Head
(
Rev. 1)
Page 30 of 34
Use a suitable JIS screwdriver to reinstall two M2x4 JIS
screws to secure the bottom of the graduated center
plate to the front frame (see
Figure 122 – Secure the plate with two screws (bottom)
Reinstall the Plastic Finger Grips
Grasp the two ocular tubes and slide them to their
maximum spacing such that the screw holes for the
plastic finger grips are not obscured by the front frame
(see
, left-hand side holes are shown).
Figure 123 – Set the ocular tubes to their maximum spacing
Place the left-hand finger grip into position over the
left-hand ocular tube (see
Figure 124 – Place the finger grip into position (left)
Use a suitable JIS screwdriver to reinstall an M2x10 JIS
screw into the front of the left-hand finger grip, to
secure it to the left-hand ocular slide (see
Figure 125 – Reinstall front screw into the finger grip (left)
Use a suitable JIS screwdriver to reinstall two M2.6x5 JIS
screws through the two holes in the left-hand ocular
slide (from the back) to secure the left-hand finger grip
to the ocular slide (see
Figure 126 – Reinstall back screws into the finger grip (left)
Place the right-hand finger grip into position over the
right-hand ocular tube (see
Figure 127 – Place the finger grip into position (right)