DMTA-10022-01EN, Rev. C, January 2015
Custom Setups for Single Element Transducers
173
Figure 10-7 Comparing pulser powers set to 60 V and 200 V
When the 45MG displays the
SAT
flag below the thickness reading, the input voltage
from the transducer is above the maximum range, and proper measurements cannot
be made. This can normally be corrected by lowering the
PULSER POWER
until the
SAT
flag no longer appears.
10.6 Time-Dependent Gain Curve
With single element transducers, the 45MG either uses the automatic gain control
(AGC) [see “Configuring Measurement Parameters” on page 141 for details on how to
activate the AGC] or the time-dependent gain (TDG) functions to automatically adjust
receiver gain to an optimum level when an echo is detected.
The 45MG offers three parameters for drawing a time-dependent gain curve:
INIT
GAIN
,
TDG SLOPE
, and
MAX GAIN
(see Figure 10-8 on page 174). From the initial
gain level, receiver gain slopes up to the maximum gain level at the rate determined
by the
TDG SLOPE
setting. When you adjust any of the received gain parameters, a
black time-dependent gain curve is displayed that clearly indicates the zones of initial
gain, slope, and maximum gain.
Initial pulse
Initial pulse
NOTE
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...