45MG — Getting Started Guide
5
2.
Couple the transducer on the thick sample of the test block.
3.
Press
[CAL VEL]
.
4.
Once the thickness reading is stable, press
[ENTER]
.
5.
Use the arrow keys to enter the known thickness.
6.
Press
[CAL ZERO]
.
7.
Place couplant on the surface of the thin sample of the test block.
8.
Couple the transducer on the thin sample of the test block.
9.
Once the thickness reading is stable, press
[ENTER].
10. Use the arrow keys to enter the known thickness.
11. Press
[MEAS]
.
Note
: For more information about calibration, and about
selecting the correct transducer for a specific application, refer to the
45MG Ultrasonic Thickness
Gage — User’s Manual
(P/N: DMTA-10022-01EN), or contact Olympus.
Keypad Functions
The 45MG comes with either an English or international keypad. The functions on both keypads are
identical. Many keys on the international keypad contain pictograms instead of text labels. In the
45MG user documentation, keypad keys are referred to using the English label, displayed in bold and
within brackets (ex.:
[FILE]
).
Each key is labelled according to its primary function. The area immediately above certain keys
indicates a secondary key function. The
[
]
,
[
]
,
[
]
,
[
]
, and
[ENTER]
keys are used to select menu
items and screen parameters, and to change parameter values. Use the
[MEAS]
key at any time to
return to the measurement screen.
Transducer on a
thick sample
Transducer on a
thin sample
English keypad
International keypad