DMTA-10004-01EN, Rev. D, November 2016
Chapter 13
224
7.
In the
SAVE SETUP
screen:
a)
If needed, in
SAVE AS
, edit the setup name.
b)
In the
SAVE TO
list, select the desired custom setup location where you want
to the save the setup.
c)
Select
SAVE
.
8.
Press
[MEAS]
to return to the measurement screen.
The saved setup becomes the active setup.
13.2 Quickly Adjusting Waveform Parameters For Single Element
Transducers
With single element transducers, you can quickly adjust individual waveform
parameters using the
[WAVE ADJ]
key.
To quickly adjust individual waveform parameters
1.
Ensure that a single element transducer is connected to the 38DL PLUS.
2.
From the measurement screen, press
[WAVE ADJ]
.
The waveform adjustment parameter appears on the measurement screen above
the thickness value (see Figure 13-2 on page 224).
Figure 13
‑
2 Adjusting the waveform parameters
Waveform parameter
value
Waveform parameter
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...