DMTA-10009-01EN [U8778346], Rev. D, November 2016
Chapter 5
68
Figure 5
‑
4 Performing the material sound velocity calibration on a 5
‑
step test block
2.
To perform the zero calibration (see Figure 5-5 on page 69):
a)
Place a drop of couplant on the surface of the thin part of the test block.
b)
Couple the transducer to the thin part of the test block, and then press
[CAL ZERO]
.
c)
When the thickness reading is stable, press
[ENTER]
.
d)
Using the arrow keys, edit the thickness value to match the known thin
thickness of the test block.
12.7 mm
(0.500 in.)