DMTA-10009-01EN [U8778346], Rev. D, November 2016
Basic Operation
69
Figure 5
‑
5 Performing the zero calibration on a 5
‑
step test block
3.
Press
[MEAS]
to complete the calibration and return to the measurement screen.
If you turn off the instrument before pressing
[MEAS]
, the velocity is not updated to
the new value; instead the instrument retains the previous value.
When the 38DL PLUS detects an error in the calibration procedure, it successively
displays the following messages in the help text bar before returning to the
measurement screen:
“
Potential wrong echo detected!
”
“
Invalid calibration results!
”
In this case, the velocity is not changed. The probable case is that an incorrect
thickness value was entered.
2.54 mm
(0.100 in.)
IMPORTANT
NOTE
Summary of Contents for 38DL PLUS
Page 6: ...DMTA 10009 01EN U8778346 Rev D November 2016 Table of Contents vi...
Page 8: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Abbreviations viii...
Page 24: ...DMTA 10009 01EN U8778346 Rev D November 2016 Important Information Please Read Before Use 16...
Page 26: ...DMTA 10009 01EN U8778346 Rev D November 2016 Introduction 18...
Page 46: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 2 38...
Page 56: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 3 48...
Page 102: ...DMTA 10009 01EN U8778346 Rev D November 2016 Appendix 94...
Page 106: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Tables 98...