Test of medical power adapter
The power supply is Class I. Protective earthed parts are only located inside the power supply
enclosure and therefore are not accessible. The accessible parts of the enclosure and output
are DOUBLE INSULATED from all mains voltage electrical parts. Ground continuality test is not
relevant.
Leakage test of the adapter is made by having the measurement equipment connected to the AC
plug, no need to open the plastic enclosure.