3.1.
Checking the installation
It is checked that, at least, all connections have been made and that the complete installation
complies with current standards in force.
CAUTION
OLDHAM is not responsible for the compliance of the complete electrical safety system.
The MX48 unit is switched on by means of circuit breakers * provided for that purpose and which
ensure protection of the mains power unit.
* The circuit breakers are to be selected according to the power consumption levels specified by the
manufacturer and the length of the electric cables.
3.2.
Switching on the unit
CAUTION
The handling operations and adjustments described in these paragraphs are strictly
reserved for authorized personnel as they are liable to affect detection safety.
To start up the MX48 unit, you must:
-
swivel the front panel,
-
press the ON/OFF button located to the bottom left-hand side of the FRONT
circuit: see Figures 3 (item A).
-
The display panel then shows, for example:
MX 48/52 V1.0
The unit then goes into INITIALIZATION mode for one minute. Consequently, all the alarms are
inhibited and the current outputs are 1 mA for the channels in service. The unit then performs a self-
test * on its buzzer and all its night-emitting diodes. At the end of this one-minute period, the
channels in service return to normal operation and the corresponding alarms and relays are enabled.