42
Table No. 2
Chlorine
C1
2
Hydrochlorid
e acid
HCl
Hydrocyanic
acid
HCN
Ammonia
NH
3
Ammonia
NH
3
Nitrogen
oxide
NO
Nitrogen
Dioxide
NO
2
Sulfur
Dioxide
SO
2
Reference
6 313 809
6 313 804
6 313 805
6 313 799
6 313 800
6 313 802
6 313 801
6 313 819
Standard range (1)
10
30
10
100
1000
300
30
30
Measurement principle
Electrochemical
Electrochemical
Electrochemical
Electrochemical
Electrochemical
Electrochemical
Electrochemical
Electrochemical
Display resolution (1)
0,1
0,1
0,1
1
1
1
1
1
Accuracy (2)
0,25
1
0,25
5
30
10
1
1
Repeatability (3)
2
2
2
2
2
1
1
1
Deviation of zero/sensitivity
(4)
0.5 / 5
0.5 / 5
0.5 / 5
1 / 2
1 / 2
0.5 / 3
0.5 / 5
0.5 / 2
Response time (5)
< 60
< 80
< 60
< 60
< 60
< 30
< 30
< 60
Temperature (6)
-20 to +40
-20 to +40
-20 to +40
-20 to +40
-20 to +40
-15 to +40
-20 to +40
-20 to +50
Relative humidity and
pressure range (7)
10 – 90 % RH
1 bar ± 20 %
15 – 95 % RH
1 bar ± 20 %
15 – 95 % RH
1 bar ± 20 %
15 – 90 % RH
1 bar ± 10 %
15 – 90 % RH
1 bar ± 10 %
20 – 90 % RH
1 bar ± 20 %
15 – 90 % RH
1 bar ± 20 %
15 – 90 % RH
1 bar ± 10 %
Service life (8)
30
24
24
24
24
30
40
30
Storage conditions (9)
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
Response time (10)
Sensors functional immediately after start up of instrument
Notes:
- Exposure to high levels of organic solvents can damage the sensors.
- Exposure to gases at higher levels than detector’s range can damage it Recalibrate sensors if they go out of range.
1 -
In ppm unless otherwise specified
.
2 -
At 50% of scale (same unit as range)
3 -
As % of signal read unless otherwise specified
4 -
Nominative values in normal use conditions per month as % of scale for zero and as % of measurement for sensitivity
5 -
In seconds at 90% of final value
6
-
In °C
7
-
Without condensation
8
-
Average noted per month 12 month guarantee
9
-
All sensors must be protected from air when stored
10
-
Time from start- up for optimal performance of sensor.