44
8.2.5.3
Obj. Heater
Press on
Obj
icon
, to enter the Objective Heater calibration submenu (see Figure 62).
Note
►
Make sure that the Objective Heater is enabled (see paragraph 8.2.3).
a
b
1
2
4
3
Figure 62. Settings
–
Calibrations - Obj. Heater (a
–
b).
At this point you can choose:
1.
By pressing
Log
button (see pointer 1 in Figure 62 b) you can select parameters obtained by calibrations
previously performed (see Figure 63 b).
a
b
Figure 63. Settings
–
Calibrations
–
Obj. Heater - Log (a
–
b).
2.
By pressing
Factory
button (see pointer 2 in Figure 62 b) UNO-CONTROLLER will recover the factory
offset.
3.
To perform the
Manual
procedure, follow the instructions below:
Note
►
Make sure that a Chamber Calibration has been performed (see paragraph 8.2.5.1).
•
Make sure that T Sensor is connected to UNO-CONTROLLER (see paragraph 7.4).
•
Secure with commercial adhesive tape the T Sensor to the bottom of the sample holder you use during
standard operation (Petri dish, Chambered Glass Slide, etc.) as shown in Figure 64. Do not cover the
tip of the T-Sensor with the adhesive tape.