34
Ambient temperature
[°C]
Desired well
temperature
[°C]
Base
Offset
[°C]
Lid
Offset
[°C]
H.M.
Offset
[°C]
23.0 ± 1.0
(Suggested ambient temperature)
32.0
-1.1
-2.2
-1.7
33.0
-1.1
-2.4
-2.0
34.0
-1.2
-2.7
-2.2
35.0
-1.3
-2.9
-2.5
36.0
-1.4
-3.2
-2.7
37.0 (Default)
-1.5
-3.5
-3.0
38.0
-1.7
-3.8
-3.2
39.0
-1.8
-4.2
-3.5
Table 3: Reading offset estimated value: extrapolated values from trend lines equations. Configuration; 35mm Petri Dish adapter, with
Humidifying Module heater. Standard values are pointed out (specimen temperature 37°C.)
Advantages of this approach:
High accuracy
Well suited for expert user
Disadvantage:
Further calibration is needed in case room temperature differs more than +/-2°C
from factory calibration temperature of 23°C.
Offset values are valid only for one Set Point temperature. The Offset values must
be experimentally re-evaluated for each different set point.
If the sample adapter changes, please modify the Offset and re-calibrate the chamber, to
ensure that the sample will be at the correct temperature.
Offset values are evaluated according to the following:
1.
Maintaining sample temperature error within ± 0.3 °C
2.
Avoiding condensation on the lid glass
3.
Improving the gas stream humidity level by adjusting the offset of HM
13
Unit Operation Through Push Button Interface
Main Page, Main Menu And Submenu Description
Figure 18. Controller Display "Main Page" and function buttons
Set Poi nt 37. 0° C
Fr ee Sensor 37. 0° C
Base 37. 0 HM 37. 0