CO2-O2 Measurement process
LEO allows measurement of the CO2 and O2 concentration in two different modes:
•
Diffusion Mode
,
without pump running. Use Diffusion Mode when the sampling gas has an
overpressure higher than 10 mbarg (0.145 psig).
Note
►
Do not exceed 300 mbar (4 psig) upstream of LEO
Note
►
LEO can operate at minimum flow rate of 30 ml/min and at maximum flow rate of 400 ml/min.
Tip
►
The optimal flow rate of the sampling gas is 150 ml/min.
•
Aspiration Mode
,
with pump running. Use Aspiration Mode when the sampling gas is at
atmospheric pressure.
Tip
►
The default value of the gas flow rate sucked by LEO is 150 ml/min. To change this value, see
paragraph 9.3.3.
To start a gas measurement process, press the
ON/OFF
button and wait until the
Warming up time
ends.
It will take 15 minutes for CO2-O2-MODULE to achieve stability. During this time, the message
“LEO is wa
rming
up”
appears (1 in Figure 45).
Figure 45 . LEO is warming up.
CO2 and O2 gas concentration measurement in Diffusion Mode cannot occur
simultaneously with a CO2
–
O2
gas concentration measurement in Aspiration Mode.
1