H301-NIKON-TI-SR
User Manual
March 2015
www.oko-lab.com
Pa
ge
12
8.
Connecting the Gas Supply and Filling the Humidity Module
A single silicon tubing carries output gas from the Okolab Gas Controller to the H301-NIKON-TI-SR. Silicon tubing con-
nects to a gas input - brass opening - located on a side of the H301-NIKON-TI-SR. See Figure 16. Connect the silicon
tubing by gently pushing it onto brass opening.
H301-NIKON-TI-SR includes an integrated humidity module in order to maintain a high humidity percentage inside the
chamber. Fill the humidity module inside the chamber through the water input - brass opening - located on a side of
the H301-NIKON-TI-SR (See Figure 16) using a silicon tubing 2 mm ID and a syringe*.
*NOTE: Do not exceed 0.05 l
Gas Input
Water Input
Figure 16. Connection with gas supply.
9.
Working with Perfusion
H301-NIKON-TI-SR features 8 perfusion holes for the insertion of perfusion tubing up to 2.5 mm in outer diameter.
Small screws plug the perfusion holes when not in use. (
Grub screws M3x3
). Remove small screws as needed before
introducing perfusion tubing.
Figure 17 shows location of perfusion holes.
1. Frontal view
2. 3D view
Perfusion dedicated opening
Figure 17. Perfusion