32
to send to your incubator/device. Press the
Total
flowmeter to set the gas flowrate in the range 0.2÷1.0 l/min using
the slider, as shown in Figure 48.
a
b
Figure 47. Flowrates View (a
–
b)
1
2
Figure 48. How to change the total flow rate.
W
hen setting the flowrate for values below 0.2 l/min, the warning message “
Total Flow not allowed
”
appears (see Figure 49).
Figure 49. Flowrates for values below 0.2 l/min.
Note
►
You can set the flowrate at 0.0 l/min, in this case the gas flowrate will be paused and output flow
will be interrupted (see Figure 50). In this working mode three dots will appear in Homepage instead of CO2%
and O2% current value and the indicator will turn yellow.