2.  When installing the remote 

probe in pipelines or vessels, 
disconnect the probe from the 
transmitter to prevent breaking 
wires in the probe assembly.  
Reconnect the probe to the 
transmitter after the probe is 
installed. *

The probes are 

not interchangeable as 
each probe has its cali-
bration data stored in an 
EEPROM on the trans-
mitter

.   

3.  Connect the transmitter to a 18 

to 28 volt DC power supply and 
turn power on. 

4. 

Wait for the dew point reading 
to stabilize. The transmitter 
range is limited by the response 
of the internal sensor. 

There is 

no over/under range indica-
tion, and the accuracy is not 
rated beyond -40°F to +80°F      
(-40°C to +20°C). 

 

RECALIBRATION 

The DPX-200 is a microprocessor 
based-instrument and has no user 
adjustments.  *

The probes are 

not interchangeable as 
each probe has its calibra-
tion data stored in an 
EEPROM on the transmitter

.  

The system can be recertified to 
NIST traceable standards by return-
ing it to Ohmic Instruments Co. 
Recertification is recommended on 
a yearly basis. 

 

OPTIONAL RS-232 INTERFACE– 
Connecting the DPX-200 to Your 
Computer 

Connecting to a serial port   If your 
DPX-200 includes the optional RS-
232 adapter cable, it can be con-
nected to your computer’s 9-pin PC 
serial (COM) port. If your computer 
has a 25-pin COM port, 25-pin to 9-
pin serial adapters are available from 
most computer stores and can be 
used with the adapter cable [Serial 
data is transmitted using Pin 5 
(ground), Pin 3 (transmit), and Pin 2 
(receive)]. Some computer COM 
ports and serial adapters are not fully 
implemented  (not all pins are used).  
If the DPX-200 is connected to a 
partially implemented serial port, 
functions may be erratic or the unit 
may fail to communicate. Consult the 
owner’s manual for your computer if 
this occurs. Set up your communica-
tions program as described on Page 
6. 

Connecting to a USB port   If your 
DPX-200 includes the optional RS-
232 adapter cable, it can be con-
nected to your computer’s USB port 
via an optional RS-232 to USB 
adapter from Ohmic. Ohmic recom-
mends using an RS-232 to USB 
adapter only with Windows XP

©

 and 

later versions. Connect the DPX-200 
using the RS-232 adapter and RS-
232 to USB adapter. Turn on your 
computer. When your computer rec-
ognizes the device click OK and fol-
low the instructions for installing the 
driver software. Set up your commu-
nications program as described on 
Page 6. 

Summary of Contents for DPX-200

Page 1: ...F SLS 554 A Phone 410 820 5111 Toll Free 800 626 7713 Fax 410 822 9633 www ohmicinstruments com Sales sales ohmicinstruments com Service service ohmicinstruments com Ohmic Instruments 3081 Elm Point I...

Page 2: ...ssion of Ohmic Instruments Information contained in this manual is believed to be accurate and reliable however no responsibility is assumed by Ohmic Instruments for its use Ohmic reserves the right t...

Page 3: ...ryer systems as well as ambient conditions The DPX 200 determines the dew point temperature from the measurements of water vapor and dry bulb temperature of the air sample These dew point calculations...

Page 4: ...smitter operates as a 4 wire device DO NOT TIE THE POWER CONNECTION TO THE SIGNAL OUTPUT ON THE TRANSMITTER THIS WILL DAMAGE THE UNIT For analog output values refer to the table on Page 9 1 Power 18 2...

Page 5: ...nnecting to a serial port If your DPX 200 includes the optional RS 232 adapter cable it can be con nected to your computer s 9 pin PC serial COM port If your computer has a 25 pin COM port 25 pin to 9...

Page 6: ...indicating the version of the embedded program The computer will then display three numbers Relative Humidity Dry Bulb Temperature and Dew Point New readings will be displayed every few seconds and a...

Page 7: ...from 0 to 32767 For n 0 1 and 2 automatic trans mission is turned off Default value is n 3 seconds Note To enable the screen to dis play what you are typing go to the menu bar in Tera Term click on Se...

Page 8: ...nication Setup HC 610 10K NTC Thermistor 45 to 90 F 40 to 80 F 4 F from 40 to 20 F 2 F from 20 to 80 F 0 200 F 1 F Probe 0 to 185 F Electronics 40 to 120 F 175 psig for remote probe 10 sec for 63 step...

Page 9: ...27 0 542 5 733 4 1 833 9 867 32 3 000 13 600 63 4 292 17 733 26 0 583 5 867 5 1 875 10 000 33 3 042 13 733 64 4 333 17 867 25 0 625 6 000 6 1 917 10 133 34 3 083 13 867 65 4 375 18 000 24 0 667 6 133...

Page 10: ...he instrument returned transportation prepaid SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT All shipments of Ohmic Instruments products should be made via United Parcel Service or Best Way prepaid...

Page 11: ...11...

Page 12: ...12...

Reviews: