GR-C-MANUAL

 (Rev. 1) 2017-11 

 

improper maintenance, repair, use and/or abuse leading to malfunction of the device, 
replacement is the sole responsibility of the user.   

 
 WARNING:  DO NOT attempt to operate this device unless you are properly trained in its use or are   
 

 

supervised by someone educated on the proper use of the apparatus (REF. CGA Pamphlet  

 

 

G-4 “Oxygen”).   

 
WARNING:  DO NOT use this device on patients who have stopped breathing without  using proper 

resuscitation equipment. 

 
TO MINIMIZE THE RISK OF EXPLOSION OR FIRE:
 

•   Make sure that the person installing or using the cylinder, Gas Regulator and other oxygen apparatus has clean  
    hands that are free of hand lotions, petroleum based product, hair spray or any other similar products. 
•   Inspect the cylinder valve and Gas Regulator thoroughly for dust, oil and grease.  DO NOT 
    USE THE CYLINDER IF OIL OR GREASE OR ANY OTHER PETROLEUM-BASED FLAMMABLE       
    SUBSTANCE IS PRESENT.  Inform your gas supplier of this condition immediately. 
•   DO NOT allow Oxygen or Oxygen equipment to be exposed to fire, heat, sparks, electrical switches or  
    other possible sources of ignition.  DO NOT store or use oxygen equipment at temperatures of less than 0

o

F  

    (-17

o

C) or more than 120

o

F (48

o

C) 

•   NO SMOKING – Remove matches, cigarettes, lighters and lighter fluids from the patient and from the   
    oxygen therapy area before administering oxygen.  Remove other flammable materials from the area.  
•   DO NOT allow cylinders to tip or fall.  Never allow the temperature of the cylinder contents to exceed     
    120

o

F (48

o

C).  DO NOT store cylinders near sources of heat or flame 

 

INTENDED USE 

                                                                                                 
The Amvex Click Style Gas Regulator is used to control the gas pressure of medical compressed gas accurately 
under the direction of a healthcare processional. The Gas Regulators are available for Nut and Nipple, Yoke 
style CGA and Bullnose fittings.  The units are machined brass finished with an anodized aluminum body. The 
Amvex Click Style Gas Regulators are available in a variety of flow rates from 0-4, 0-8, 0-15, 0-25 L/min. 

 
SPECIFICATIONS 
 

Gas Regulator Description 

1.

 

Yoke style (used on small D and E size cylinders). This connection must have a sealing washer. 

2.

 

Nut and Nipple style. There is no sealing washer used with this connection.  Tighten this fitting to the  
cylinder valve with a wrench. 

 

Flow Range 

Accuracy 

0 - 1 L/min 

+/- 20% from 0-1 L/min 

1.5 - 25 L/min 

+/- 10% above 1 L/min 

Inlet pressure max 3000 PSI maximum. Preset 50 PSI +/-10 PSI

 

                                                                                

Summary of Contents for AMVEX CLICK STYLE

Page 1: ...TORS CLICK STYLE AND MINI CLICK STYLE Rx Only Basic matrix shown Consult the Ohio Medical website for full matrix or contact your Ohio Medical representative The product package label contains your product identifier and description ...

Page 2: ...anufacturer Catalog Number Device Identifier Rx Only Caution Federal USA Law restricts this device to sale by or on the order of a licensed healthcare provider RECEIVING INSPECTION Remove product from package and inspect for damage Verify that the model received is in working order If product is damaged or incorrect do not use Contact your dealer equipment provider or manufacturer ATTENTION It is ...

Page 3: ...OT store or use oxygen equipment at temperatures of less than 0o F 17o C or more than 120o F 48o C NO SMOKING Remove matches cigarettes lighters and lighter fluids from the patient and from the oxygen therapy area before administering oxygen Remove other flammable materials from the area DO NOT allow cylinders to tip or fall Never allow the temperature of the cylinder contents to exceed 120o F 48o...

Page 4: ... a wrench c If the Gas Regulator has a Bullnose or DIN inlet nut 477 connection install the inlet nut on the cylinder valve and hand tighten ATTENTION Sealing washers should be checked for damage or contamination before each use If either is visible replace immediately It is recommended that a new seal be used each time the cylinder is changed ATTENTION Fittings attached to auxiliary ports are fac...

Page 5: ...low from the outlet orifice assembly until the cylinder valve is closed Reference CGA Pamphlet P 14 Accident Prevention in Oxygen Rich and Oxygen Deficient Atmospheres It is very important to always close the cylinder valve when not in use Removing The Regulator From A Cylinder 1 Close the cylinder valve 2 Wait for the oxygen to flow out of the Gas Regulator then turn the knob to the OFF position ...

Page 6: ...pments or failures resulting from accidents misuse abuse neglect mishandling alteration misapplication or damage that may be attributable to acts of God The Company DOES NOT HONOR VERBAL STATEMENTS CONCERNING THE WARRANTY The distributor and or dealer are not sanctioned to create verbal warranties about the product described in this agreement Any statements will not be honored or be made part of t...