background image

Operator/Maintenance Manual

®

550102 (Rev. 12) 10/2021

Multipurpose Suction Unit

Medical Equipment

Summary of Contents for 756000

Page 1: ...Operator Maintenance Manual 550102 Rev 12 10 2021 Multipurpose Suction Unit Medical Equipment...

Page 2: ...rsons properly trained in medical suction techniques and in the operations of suction equipment Improper use could cause injury Thoroughly read this operations manual to familiarize yourself with the...

Page 3: ...ipment can affect medical electrical equipment This product is intended for use in the electromagnetic environments specified The end user of this product should assure it is used in such an environme...

Page 4: ...s in good working order and that no damage has occurred during shipment SET UP INSTRUCTIONS 1 Visually inspect all components for physical damage that may have occurred during shipping 2 After removin...

Page 5: ...patient use The Instavac mobile aspirator is now ready to place into service OPERATING INSTRUCTIONS PRE USE CHECKOUT PROCEDURE WARNING The Pre Use Checkout Procedure must be performed before using the...

Page 6: ...DE SELECTION switch Plug the vacuum port located at the top of the back pole and adjust the vacuum to the desired vacuum level by using the VACUUM REGULATOR knob 4 Switch the unit to INTERMITTENT suct...

Page 7: ...cessary vacuum the filter or remove and rinse in water After the filter is completely dry reinstall the filter 4 Inspect the overflow trap assembly for any evidence of an accidental overflow If an ove...

Page 8: ...to service Follow steps in section 10 to verify and adjust the device settings as necessary PREVENTATIVE MAINTENANCE The Instavac is warranted for its intended life of three 3 years Performing period...

Page 9: ...Apply a light coat of a silicone based lubricant i e Dow Corning 111 to the O rings and reassemble back onto the black cartridge 7 Reassemble the vacuum regulator and mount the regulator in the Instav...

Page 10: ...internal diaphragm for the check valve is facing AWAY from the solenoid manifold 7 Proceed to Section 10 Performance Test Adjustment to set vacuum flow CAUTION Avoid aspirating moisture into the pump...

Page 11: ...ment 10 Performance Test Adjustment After performing any service or repair of the Instavac including preventative maintenance and cleaning for overflow follow the steps below to verify that the device...

Page 12: ...ontrol panel to full vacuum 2 With a 1 16 Allen wrench loosen the set screw Turn the flow adjustment valve until the flow specification section 12 is reached Tighten the set screw 11 Troubleshooting C...

Page 13: ...Problem High flow on INTERMITTENT models 1 Flow adjustment valve for intermittent flow is set too high 1 Adjust the flow rate according to the directions found on page 12 2 Faulty 2 way solenoid on th...

Page 14: ...apacitor replaces 756370 Capacitor 10 F Terminal Cover Mounting Strap 756608 Kit Pump Replacement 120V replaces AI5511 Vacuum Pump Gast Capacitor Fittings Wire Terminals Instructions 756605 Kit Vane R...

Page 15: ...550102 Rev 12 10 2021 15 Figure 1 Instavac I 756000 756010 See page 18 for Parts List 14 Detailed Drawings Parts List...

Page 16: ...16 550102 Rev 12 10 2021 Figure 2 Instavac I 756000 756010 See page 18 for Parts List...

Page 17: ...550102 Rev 12 10 2021 17 Figure 3 Instavac I 756000 756010 See page 18 for Parts List...

Page 18: ...Connector Business Machine 19 1 703010 Fitting 1 8 NPT Male x 5 16 ID Hose Barb 20 1 AI5657R Hex Nipple 21 1 756354 Mount Standoff Wire with Lock 22 1 756260 Kit Fan 120V 23 2 AI5526 01 Muffler Exhau...

Page 19: ...550102 Rev 12 10 2021 19 Figure 4 Instavac C 757000 757010 See page 22 for Parts List...

Page 20: ...20 550102 Rev 12 10 2021 Figure 5 Instavac C 757000 757010 See page 22 for Parts List...

Page 21: ...550102 Rev 12 10 2021 21 Figure 6 Instavac C 757000 757010 See page 22 for Parts List...

Page 22: ...Fitting 1 8 NPT Male x 5 16 ID Hose Barb 18 1 AI5657R Hex Nipple 19 1 756354 Mount Standoff Wire with Lock 20 1 756260 Kit Fan 120V 21 2 AI5526 01 Muffler Exhaust Bottle 22 8 756353 Bumper Vibration 3...

Page 23: ...550102 Rev 12 10 2021 23 Notes...

Page 24: ...information is under license from Ohio Medical LLC Any use other than that authorized by Ohio Medical LLC is prohibited Ohio Medical and the Ohio Medical logo are registered trademarks of Ohio Medical...