14
3.6 Parts Counting
The Voyager balance can be set to either an Easy Count or Advanced Count parts counting method.
Easy Count
setting enables a simplified method for counting parts. When selected, Easy Count displays a screen
which requires that the number of sample parts be entered. After this entry, putting a quantity of samples on the pan,
the balance will display the actual number of samples. Since the balance determines the quantity based on the
average weight of a single part, all parts must be reasonably uniform in weight.
Advanced Count
setting contains a number of entry screens which include assigning a library name, filling and
sorting applications and statisical information which can be printed.
3.6.1 Easy Count
Go Back
Help
Enter
Procedure
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CHANGE MODE.
• Press
Enter
button.
• Press
or
arrow button and select PARTS
COUNTING.
• Press
Enter
button.
• Press
or
arrow button and select EASY
COUNT.
• Press
Enter
button.
• Enter the numeric sample size as indicated on the
screen using the arrow buttons to set the value.
• Press
Enter
button.
• Place sample size on pan and press
Enter
button.
• Remove samples from pan and place samples to be
measured on the pan. The display indicates the num-
ber of pieces based on the sample size. Repeated
batches of samples may be placed on the pan and
counted. Sample shown below indicates 10 pieces.
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CALIBRATION
CUSTOM MENU
CHANGE MODE
0.00
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
MAIN MENU
CHANGE UNITSPART
S
ET BALANCE
RETURN TO
WEIGH
CALIBRATION
CUSTOM MENU
CHANGE MODE
CHANGE MODE
BASIC WEIGHING
PARTS COUNTING
FILLING
ANIMAL WEIGHING
CHECK WEIGHING
DIFF WEIGHING
FORMULATION
QUICK CHECK
STATISTICS
DENSITY
MAIN MENU
CHANGE UNITSPART
S
ET BALANCE
RETURN TO
WEIGH
CALIBRATION
CUSTOM MENU
CHANGE MODE
CHANGE MODE
BASIC WEIGHING
PARTS COUNTING
FILLING
ANIMAL WEIGHING
CHECK WEIGHING
DIFF WEIGHING
FORMULATION
QUICK CHECK
STATISTICS
DENSITY
PARTS COUNTING
EASY COUNT
ADVANCED COUNT
RECALL
EXIT TO WEIGH
EASY COUNT
0%
100%
PIECES
UNSTABLE
OPTIMIZE
MAIN
A.P.W. =
SIZE =
TARE WT =
WEIGHT =
CONTRAST
NEW
ENTER SAMPLE SIZE
10
<UP>/<DOWN> TO SCROLL CHARACTERS
<LEFT>/<RIGHT> TO MOVE CURSOR
<ENTER> TO ACCEPT
<BACK> TO ABORT
EASY COUNT
0%
100%
PIECES
UNSTABLE
OPTIMIZE
MAIN
A.P.W. =
SIZE =
TARE WT =
WEIGHT =
CONTRAST
NEW
SET A.P.W.
PLEASE PUT:
10
SAMPLES
AND PRESS < ENTER > . . .
EASY COUNT
0%
100%
PIECES
UNSTABLE
OPTIMIZE
MAIN
A.P.W. = 30.00
SIZE = 10
TARE WT = 0.00
WEIGHT = 300.02
CONTRAST
NEW
10
Summary of Contents for Voyager V00640
Page 61: ...55...
Page 62: ...56...
Page 65: ...1 Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 Operating instructions balances...
Page 154: ...Operating instructions balances O T O T O T O T OHAUS...
Page 243: ......
Page 245: ......