EN-6 Valor 3000 Series Scales
4. SETTINGS
The Menu allows the user to customize scale settings.
4.1 Menu Navigation
Press and hold
MODE/UNIT/Menu
button until SEtUP is displayed, then release it.
Press
Yes
to enter the displayed menu. or press
No
to advance to the next menu.
Once in the menu, press
Yes
to view the menu item setting or press
No
to move to
the next menu item. When viewing the setting, press
Yes
to accept the setting, or
press
No
to change the setting. When End is displayed, press
Yes
to advance to the
next menu or
No
to return to the top of the current menu.
4.2 Setup Menu
STABILITY RANGE
Set the stability range to 0.5d,
1d
, 2d or 5d (divisions).
AUTO ZERO TRACKING RANGE
Set the automatic zero tracking range to OFF,
0.5d
, 1d, 3d
(divisions).
FILTER LEVEL
Set the filter level to LO (low),
ME
(medium), or HI (high).
BACK LIGHT
Set the back light functionality to OFF (always
off), ON (always on) or
Auto
(turns on when a
key is pressed or the displayed weight changes).
AUTO OFF
Set the automatic shut off functionality to
ON
(shut off after
5 minutes) or OFF (stays on).
POWER ON UNIT
Set the unit of measure displayed at startup to g, kg, lb, oz
or lb:oz.
GLOBAL RESET
Select YES (reset menus to factory defaults) or NO (do not
reset). Units will be set to factory defaults.