2-17
CHAPTER 2 DIAGNOSIS
TABLE 2-14. TROOPER COUNT SCALE ERROR CODES (Cont.).
CAUSE
Capacity point parameter
check failed during program
running.
Incorrect calibration weight
was used.
keypad input number is over
limit.
Scale is not stable at power
on.
Interference between plat-
form and load cell.
REMEDY
Recalibrate scale. If the problem per-
sists, replace the Main PC board or Load
Cell.
Recalibrate the scale.
Enter proper value within limits.
Check that the surface the scale is rest-
ing on is stable.
Check that proper clearances exist be-
tween the cover and the load cell under
the platform.
SYMPTOM
[CAP][data][Error]
[Load][Error][ ]
[Entry][Too][High]
[Start][Stable][Error]