12
4
Setup
4.1
Set temperature unit
To set the temperature, you need go into the setup screen. When
Press and hold button-
Mode/Setup
until the setup icon
appears on the display and the
current temperature unit blinks.
Press button-
up
or button-
down
to switch between ºC and ºF .
Press button-
Read
to confirm your selection.
Then
Continue with MTC temperature setting or
Press button-
Exit
to return to the measurement screen.
4.2
Set manual temperature compensation
After confirming the temperature unit, the MTC (
M
anual
T
emperature
C
ompensation)
temperature value appears on the display. Then use button-
up
or button-
down
to increase or
decrease the value to the temperature of your sample. Press button-
Read
to confirm the setting or
press button-
Exit
to return to the measurement screen.
The default MTC temperature value setting is 25 °C (77 ºF).
Note:
ºC = 5/9 (ºF -32)
4.3
Selecting a predefined buffer group
After confirming the MTC compensation temperature value, it goes to the buffer group selection.
Use button-
up
or button-
down
to select a buffer group among 4 buffer groups (3.1.1). Press
button-
Read
to confirm the setting or press button-
Exit
to leave to return to the measurement
screen.
The default buffer group is
b1
.
b1
1.68
4.01
7.00
10.01