SCOUT
®
SERIES BALANCES - STX
EN-19
4.5.4 Measuring the Density of Porous Material Using Oil
To enable this feature, enter the Item Settings
screen, and select the following:
Density Type:
Solid
Auxiliary Liquid:
Water
Porous Material:
On
To adjust the Water Temperature or Oil
Density values, touch the
Water Temp
or
Oil
Density
button.
The balance calculates water density based on
the water temperature value entered (look-up
table).
Measure the actual water temperature using a
precision thermometer.
Numeric input windows appear.
Key in the desired value, then press
Save
.
To return to the Density Determination home
screen, touch
Exit.
Press
Start
to begin density determination.
Step 1 of 3
– Weigh the Un-Oiled Sample in
Air.
Follow the screen instructions then press
Accept
to store the dry sample weight (in air).