3. After confirmation by pressing
Yes
, the message
"Clear APW?" will appear on the screen.
4. If the APW of the last Parts Counting operation needs
to be kept, press
No
when the message "Clear
APW?" displays.
5. Press
Yes
, and the message "Sample size 10" will
display with the numeral "10" (default) flashing.
6. Confirm the sample size by pressing
Yes
, and place
10 samples on the pan to display the weight. Press
No
or
Back
to increase or decrease the value as
desired.
7. Press the
Function / Mode
button so that the weight
of the 10 samples is used to establish the average
piece weight (APW). The display will show 10 pieces.
8. To view the piece weight or the total weight, press the
Function / Mode
button.
9. Place additional objects on the pan, and the
corresponding number of pieces will display.
Item Settings
Sample:
The sample size ranges from 1 to 1000. The default
value is 10.
Note:
To ensure accurate counting, the minimum piece
weight should be no less than 0.1d.