APPENDIX B THE SERVICE MENU
Navigator Series Service Manual
41
Ohaus Corporation www.ohaus.com
B.6 FILTER SETTINGS
There are four filter settings.
L1
High filtering
The display will change slower as the scale analyzes more data before
making display changes. This is used in less than ideal conditions where
the scale display is changing too often.
L2
Default
This factory setting that is useful for most situations.
L3
Lower filtering The display will update quicker than the factory default setting.
L4
Low filtering
The display will update rapidly. This setting is only useful where the
weighing conditions are optimal and when fast response is needed. The
stability indication will have a difficult time indicating a stable weight at this
setting.
In the Service Menu, press
No
until
FiLter
appears.
Press
Yes
to enter the
Filter
setting menu.
FiLtEr
The current filter setting is displayed. Press
No
to increment the setting.
L2
Press
Yes
to accept the displayed value and
advance to the next menu item,
Stability
.
L3
B.7 STABILITY SETTING
There are four stability settings.
0.5d This is setting requires the weight reading to vary less than 0.5d for the stable indicator
to be on. This setting will require more time for the stable indication. Less than ideal
weighing conditions may prevent stable indication.
1d
This factory setting that is useful for most situations.
2d
The stable indication will come on quicker than the factory default setting.
5d
This setting is useful where the weighing conditions are less than optimal and when
fast stable indications are needed.
In the Service Menu, press
No
until
Stab
displays.
Press
Yes
to enter the stability setting menu.
stab
The current stability setting is displayed.
set1d
Press
No
to increment the setting. Press
Yes
to accept the displayed value and advance to
the next menu item,
A-Zero
.
Set.5d
B.8 AUTO ZERO SETTING