MB35
EN-9
3.3 Temperature Calibration
Note: Temperature Calibration Kit is required.
ANALYZER SETUP
LOCK OUT
WEIGHT CAL
TEMP CAL
LANGUAGE
BEEPER
TIME-DATE
TEMP CAL
REMOVE PAN SUPPORT
ABORT CALIBRATION
/
TEMP CAL
INSERT CAL FIXTURE
CLOSE COVER
ABORT CALIBRATION
TEMP CAL
CALIBRATION READY
PRESS TARE/ACCEPT
ABORT CALIBRATION
TEMP CAL 100°C
CURRENT TEMP:
¾
C
TM TO CAL POINT: 15:00 MIN
ADJ CAL READING:
100
°
C
ACCEPT NEW CAL
ABORT CALIBRATION
15
TEMP CAL 100°C
CURRENT TEMP: 100
°
C
TM TO CAL POINT: 15:00 MIN
ADJ CAL READING:
100
°
C
ACCEPT NEW CAL
ABORT CALIBRATION
Adjust to thermometer reading
/
TEMP CAL 100°C
CURRENT TEMP: 100
°
C
TM TO CAL POINT: 15:00 MIN
ADJ CAL READING:
104
°
C
ACCEPT NEW CAL
ABORT CALIBRATION
ABORT CALIBRATION
TEMP CAL 100°C
CURRENT TEMP: 100
°
C
TM TO CAL POINT: 15:00 MIN
ADJ CAL READING: 104
°
C
ACCEPT NEW CAL
Adjust to thermometer reading
/
TEMP CAL 160
°
C
ABORT CALIBRATION
CURRENT TEMP: 160
°
C
TM TO CAL POINT: 0:00 MIN
ADJ CAL READING:
163
°
C
ACCEPT NEW CAL
15
TEMP CAL 160
°
C
ABORT CALIBRATION
CURRENT TEMP: 160
°
C
TM TO CAL POINT: 0:00 MIN
ADJ CAL READING:
160
°
C
ACCEPT NEW CAL
TEMP CAL 160
°
C
ABORT CALIBRATION
CURRENT TEMP: 160
°
C
TM TO CAL POINT: 0:00 MIN
ADJ CAL READING:
163
°
C
ACCEPT NEW CAL
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
Open the cover, allow unit to cool, remove the calibration fixture,
replace the pan support and pan.