APPENDIX B SOFTWARE SERVICE TOOL INSTRUCTIONS
MB90 MB120 Series Service Manual
78
Ohaus Corporation www.ohaus.com
4. Key in scale command and press 'Send'. The right side window will show the response
to the command.
5. MB90 and MB120 support the below print command.
B.4.1 Ramp ON
Select 'Ramp ON' the right hand window will show the A to D conversion signal from Load cell
to Main PCB. Press again to terminate the function.
The Ramp Value is not important as long as the value is stable and responsive to load on the
sample pan as shown below.
Stable Ramp value
Responsive Ramp value when
load indicated on weighing pan