APPENDIX A STANDARD CALIBRATION
MB90 MB120 Series Service Manual
64
Ohaus Corporation www.ohaus.com
A.2 Temperature Calibration.
You must have a temperature calibration kit to perform this procedure. If the Moisture Analyzer has
been recently used, allow at least 30 minutes before performing calibration.
Note:
The temperature calibration kit (11113857) is available as optional equipment.
Press Setup button to access the calibration menu.
Press Back to return to Home screen.
Press Weight and temperature calibration button.
Press Back to return to previous screen.