EN-8
Sapphire JS10 and JS500
Auto Shut-off
When Auto Shut-off is activated, the balance will shut off after three minutes of
non use.
1. When the balance displays
, press
On
to view the current setting On
or Off.
2. To change the displayed setting, press
Off.
3. To accept the displayed setting, press
On
. The display will again indicate
.
4. Change to another menu item, or exit as described above.
Activating Units
1. Start with the balance OFF. Press
On
until
is displayed.
2. Press
Mode
until
is displayed.
3. With
displayed, press
On
to display the grams indicator "g" with its
current setting of
g
or
g
.
4. To change the displayed setting, press
Mode
.
5. To accept the displayed setting, press
On
. The display will advance to the
next weighing unit.
6. Repeat steps 4 and 5 for each weighing unit and parts counting. When the
last unit has been accepted, the display will again indicate
.
7. Repeatedly press
Mode
to change to another menu item or to exit, press
Mode
until
is displayed, then press
On
to store settings.
Linearity Calibration
Calibration masses are available as accessories for performing this procedure.
Balance
Required Calibration Masses
JS10
5g, 10g
JS500
200g, 400g
2 Text new.pmd
6/10/2004, 9:53 AM
8