Ionizer 100A EN-5
3.2 Operating the Ionizer
Step 1. Turn on the ionizer by pressing the power switch.
Step 2. Place the sample in front of the ionizer in the working area as shown below. The
distance between the sample and the ionizer should be around 10~40cm.
Figure 3-1. Working area
Step 3. After about 5 seconds, remove the sample.
Step 4. Start weighing the sample.
40cm 5s
10cm <1s