TROUBLESHOOTING
The following table lists common problems and possible causes and remedies. If
the problem persists, contact OHAUS or your authorized dealer.
Error*
Cause of Error
How to Fix
Unit fails to
power on
Missing or blown fuse
Add or replace fuse as necessary.
E1
Plate RTD open
Not fixable by user, please contact
Ohaus.
E2
Plate RTD short
Not fixable by user, please contact
Ohaus.
E3
No stirring motion /
cannot reach speed
Not fixable by user, please contact
Ohaus.
E4
Probe RTD open
(Removing the probe
while the unit is
heating)
Switch unit to standby, then return
to normal operating mode.
E5
Probe RTD short
(Malfunctioning probe)
Switch unit to standby, remove the
probe from the unit, then return to
normal operating mode.
E6
A/D lock error
Not fixable by user, please contact
Ohaus.
E7
User Probe Error
(Plugging the probe into
the unit while it is
heating)
Switch unit to standby, then return
to normal operating mode.
E8
Plate over temperature
Not fixable by user, please contact
Ohaus.
E9
Plate under
temperature
Not fixable by user, please contact
Ohaus.
E10
Triac fault
Not fixable by user, please contact
Ohaus.
*Note
: Error code instances will stop equipment operation by default.