CHAPTER 2 - DIAGNOSTIC GUIDE
Explorer
®
Series Service Manual
19
Ohaus Corporation www.ohaus.com
IDNR error
Type data error
Contact the authorized dealer
--operation not
allowed--
There is no sample on the pan when
doing Tare
Place sample on the pan and redo Tare
Shortcut buttons
are gray and no
action when
pressing them.
The current application is in procedure. Cancel the procedure first.
Does not retain
the correct time
and date.
Battery backup for the real time clock
has lost its charge.
Replace the real time clock battery on the
Base Main PCBA, see 3.4.2
Note:
There is 40-second timeout control for print under stable requirement. If the unstable
condition continues over 40 seconds, balance will send “ES” to the printer.