RS232 Interface
EN-3
SETUP
Upon installation, the print sub-menu will be visible.
Reset
No
, Yes
Baud Rate:
300, 600, 1200, 2400, 4800,
9600
, 19200
Parity:
7 Even, 7 Odd, 7 None,
8 None
Stop Bit
1
or 2
Handshake:
Off
, XON/XOFF
Stable Only
Off
, On
Auto Print
Off
On Stable (-> Load, Load and Zero),
Interval (-> 1…3600),
Continuous
Content
Result (->Off,
On
, NUM)
Gross (->
Off
, On)
Net (->
Off
, On)
Tare (->
Off
, On)
Unit (->
Off
, On)
Mode (->
Off
, On)
Info (->
Off
, On)
Layout Format
(->Multiple, Single) Feed (->Line Feed, 4 Line Feed, Form Feed)
End Print
Exit PRINT menu
Note:
For more detailed information, see section 3.8 in the T24PE instruction manual.
OPERATION
Output
With the interface installed the scale will operate in one of the following two ways:
Auto Print = off, Stable = on or off
Pressing PRINT will send the display data. If Stable is set to On, the balance will wait for a
stable reading before sending data.
Auto Print = not off, Stable = on or off
The scale will automatically send data based on the settings in the Menu. If Stable is set to On,
only stable will be sent.
Data sent is in standard ASCII format terminated with a [CRLF] (carriage return-line feed).
[weight]
11 characters
[space]
1 character
[unit]
5 characters max (left justified)
[space]
1 character
[stability]
1 character
“?” when unstable, blank when stable
[space]
1 character
[G/N]
1 character
[Term. Char(s)]