14
Defender
™ Series Bases Service Manual
3.3
Overload Stop Adjustment
The Overload Stop gaps must be checked and reset if the Load Cell is replaced.
Figure 3-1. Overload Stops on Defender series Base
1. Remove the Platform and loosen the Jam Nuts.
2. Use the appropriate size feeler gauge in the Gap (see Table 3-1), and turn the screws until a
slight drag is felt on the feeler gauge.
3. Tighten the Jam Nut and re-check the Gap.
4. Re-adjust if necessary.
5. Cover the Platform and check for full capacity. (See Table 3-1 for Gap settings.)
TABLE 3-1. OVERLOAD GAP SETTINGS FOR DEFENDER SERIES BASES
Base Model
Pan size
Overload Stop
Center (mm)
Overload
Stop Corner
(mm)
Shipping
Spacer
D2WQS
254 x 254
10" x 10"
0.8
2
Y
D5WQS
1.5
2
Y
D12WQS
2.5
3.5
Y
D6RQDR
305 x 305
12'' x 12''
0.8
1
N
D12RQR
1
1
N
D15RQDR
1.2
1.5
N
D25RQR
1
2
Y
D30RQR
1.5
2
Y
D50RQR
2.5
3.5
Y
D6WQDR
1
1.5
N
D12WQR
0.8
1.5
N
D15WQDR
0.6
2.5
Y
D15WQR
0.6
2.5
Y
D25WQR
0.8
2
Y
D30WQR
1
2
Y
D50WQR
1
3.5
Y
D12RTR
305 x 355
12'' x 14'
1
1.5
N
D25RTR
1.5
2
Y
D60RTR
2.5
3.5
Y
D50RTR
2.5
3.5
N