EN-2 Defender Series Base
TABLE 1. LOAD CELL CONNECTION
Type1. For model i-D3K1S, i-D6K1S, i-D15K1R, i-D30K1R, i-D60K1L bases
FUNCTION
WIRE COLOR
+ Excitation
Blue
- Excitation
Black
+ Signal
White
- Signal
Red
+ Sense
Green
- Sense
Grey
Shield
Yellow
Type2. For model i-D150K1L base
FUNCTION
WIRE COLOR
+ Excitation
Green
- Excitation
Black
+ Signal
White
- Signal
Red
+ Sense
Light Yellow
- Sense
Blue
Shield
Deep Yellow
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid
locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat.
Level the base by adjusting the four leveling feet until the bubble in the level indicator is centered. A wrench may
be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts up
against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
CORRECT INCORRECT
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build up.
A damp cloth with water and a mild detergent may be used to wipe clean the external surfaces
– do not
use acids, alkalis, strong solvents or abrasive materials.
To maintain surface integrity, do not allow debris to remain on the surface of the painted or stainless steel base
frame.
Remove the pan and use fresh water to clean the base frame surface of any debris. Dry the base with a
clean cloth or allow to air dry before replacing the pan. Repeat daily for best care.
Summary of Contents for Defender 6000 Series
Page 1: ...DefenderTM 6000 Series Base Instruction Manual EN ES FR DE IT PT SE NL DK PL CZ HU...
Page 2: ......
Page 8: ......
Page 14: ......
Page 20: ......
Page 26: ......
Page 32: ......
Page 38: ......
Page 44: ......
Page 50: ......
Page 56: ......
Page 62: ......
Page 68: ......
Page 74: ......
Page 75: ......