CW-11 Indicator
Press
G/N/T/
MENU
,
, ,
,
, F xx is displayed..... This is full
scale capacity selections. xx= value last set.
Selections are:
5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120,
150, 200, 250, 300, 400, 500, 600, 750, 1000,
1200, 1500, 2000, 2500, 3000, 5000, 7500,
10000, 20000 (lb or kg).
Press
PRINT/
UNITS
button until desired capacity
value is reached.
Press
G/N/T/
MENU b
utton, Gd0.01 is displayed.
This is the graduation size. Refer to paragraph
2.8.3 Load Cell Capacity Information table. For
available selections, press
PRINT/
UNITS
button until
desired graduation value is reached.
Press
G/N/T/
MENU
button, CP 30 kg is displayed.
This is the full scale calibration point setting. The
range is from 20% to 100% Full scale capacity.
Press
PRINT/
UNITS
button until desired calibration
value is reached.
Press
G/N/T/
MENU
button, AS 3 is displayed. This
is the duration of animal sample time which can be
set from 1 to 10 seconds.
Press
PRINT/
UNITS
button until desired sample time
value is reached. See Note below.
Press
G/N/T/
MENU
button, Ad 4 is displayed. This
is the duration of animal weighing display time
which can be set from 1 to 10 seconds. The default
is 10 seconds. See Note below.
Press
PRINT/
UNITS
button until desired sample time
value is reached. See Note below.
Press
G/N/T/
MENU
button to end this block, SAVE is
displayed.
Press
G/N/T/
MENU
button, rEAD is displayed which
is the next menu or press
PRINT/
UNITS
button to
return to Setup menu.
The Indicator is now matched up with the scale
base and the Indicator parameters may now be set
and calibrated.
2.8 Initial Setup (Cont.)
2.8.4 Setup Menu (Cont.)
NOTE
: Not available if LFT is ON. Only available if ALT
is set to ANI.