EN-11
Valor 2000 Series Scales
6.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if
necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean
the housing or control panels.
6.2 Troubleshooting
The following table lists common problems and possible causes and remedies. If
the problem persists, contact Ohaus or your authorized dealer.
TABLE 6-1. TROUBLESHOOTING.
Symptom Possible
Cause(s)
Remedy
Scale will not turn on
No power to scale
Battery power used up
Verify connections and voltage.
Connect power and charge the battery
Poor accuracy
Improper calibration • Unstable
environment • Internal pressure
build-up
Perform calibration • Move scale to
suitable location • Relieve internal
pressure by loosening rear leveling feet
Unable to calibrate
LFT Menu set to ON
Unstable environment
Incorrect calibration mass
Set LFT menu to OFF (see section 5.1)
Move the scale to suitable location
Use correct calibration mass
Cannot display weight in
desired weighing unit
Weighing unit not enabled
Enter menu and enable unit
Cannot change menu
settings
LFT Menu set to ON
Set LFT menu to OFF (see section 5.1)
Scale displays Lo bAt
Battery is discharged
Connect power and charge the battery
Battery fails to charge
fully
Battery is defective
Have battery replaced by authorized
service dealer
Err 1
Weight on pan exceeds capacity
Remove weight from the pan
Err 2
Underload limit exceeded • Pan
was removed prior to power on
Install pan and re-zero
Err 9, Err 21
Factory data corrupted
Contact Ohaus or an authorized service
agent