background image

8

Positions  A,  B,  C,  and  D  are  centered  at  the  each

quarter  of  the  scale  platform.  The  following  table

shows the tolerance in “d” (division) for the shift test.

Scale

Capacity

Test

Weight

Acceptable

Tolerance

(New Scale)

Maintenance

Tolerance

(In Service)

3,000d

1,000d

"

1.0d

"

2.0d

Table No. 2 Tolerance Table for Shift Test

Factory Setup

When Champ II bench scales are connected to

indicators in the factory, a full span calibration is

conducted during the configuration. Because of

transportation, local environment and variation in the

earth's gravitation field, the original factory settings

must be verified and checked in the field before use.

Parts list

OHAUS PART NO.

DESCRIPTION

71129642

Indicator Bracket Assembly

71125105

Stainless Steel Column, 350 mm high

71125103

Stainless Steel Column, 680 mm high

71129064

Column Bracket

71123975

Load Cell, MT1241-50kg  CH15R

71123976

Load Cell, MT1241-100kg CH30R, CH60R, CH60L

71123979

Load Cell, MT1241-250kg CH100R

71124310

Load Cell, MT1260-500kg CH150R, CH300R

71102504

Load Cell Mounting Bolts, M6 x 25

71102512

Load Cell Mounting Bolts, M8 x 230

71129104

Stainless Steel Platform, 305 x 355

71129105

Stainless Steel Platform, 400 x 500

71129106

Stainless Steel Platform, 420 x 550

71125079

Platter Rubber Supporter

71125924

Corner Rubber Supporter, 305 x 355

71125006

Corner Rubber Supporter, 400 x 500 & 420 x 550

71127181

Cover for M6 Screw

71128342

Cover for M8 Screw

71125009

Adjustable Foot (All)

Summary of Contents for CHAMP II BASE

Page 1: ...Ohaus Corporation 29 Hanover Road P O Box 900 Florham Park NJ 07932 0900 Phone 973 377 9000 Fax 973 593 0359 CHAMP II SCALE with CD11 Operating and Installation Instructions...

Page 2: ...9 336 92 31 93 68 Compatibilit elettromagnetica EC1010 1 EN60950 1992 Safety Regulations IEC1010 1 EN60950 1992 Sicherheitsbestimmungen IEC1010 1 EN60950 1992 Consignes de s curit IEC1010 1 EN60950 19...

Page 3: ...MFUL INTERFERENCE IN WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM DIGITA...

Page 4: ...Capacity Ultimate Overload 300 of Full Capacity Operation Environment Champ II is designed to meet NTEP 3000d requirements to operate in a temperature range from 10 C to 40 C 0 to 95 relative humidity...

Page 5: ...REMOVE THE RED PLASTIC SHIPPING TABS IN THE CORNERS OF THE SCALE BASE Troubleshooting If operational difficulties are encountered first obtain as much information as possible regarding the problem Fai...

Page 6: ...ng a torque wrench tighten the hex bolts to the following torque specifications Model Torque CH15R 20 N m 15 Ft Ib CH30R 20 N m 15 Ft Ib CH60R 20 N m 15 Ft Ib CH60L 20 N m 15 Ft Ib CH100R 30 N m 22 Ft...

Page 7: ...platform weigh within tolerance If the scale does not pass the shift test verify the overload stop gaps before replacing the load cell No adjustment for the shift is possible If the shift test cannot...

Page 8: ...he field before use Parts list OHAUS PART NO DESCRIPTION 71129642 Indicator Bracket Assembly 71125105 Stainless Steel Column 350 mm high 71125103 Stainless Steel Column 680 mm high 71129064 Column Bra...

Page 9: ...9...

Page 10: ...e switches Power 9 VDC 150 mA AC adapter or six alkaline C type batteries Typical battery life 250 hours 100 hours continues use Weighing units g kg lb Modes Weighing parts counting Zeroing Range 2 18...

Page 11: ...enetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warranty registration card the warranty period shall begin on the d...

Reviews: