8
Positions A, B, C, and D are centered at the each
quarter of the scale platform. The following table
shows the tolerance in d (division) for the shift test.
Scale
Capacity
Test
Weight
Acceptable
Tolerance
(New Scale)
Maintenance
Tolerance
(In Service)
3,000d
1,000d
"
1.0d
"
2.0d
Table No. 2 Tolerance Table for Shift Test
Factory Setup
When Champ II bench scales are connected to
indicators in the factory, a full span calibration is
conducted during the configuration. Because of
transportation, local environment and variation in the
earth's gravitation field, the original factory settings
must be verified and checked in the field before use.
Parts list
OHAUS PART NO.
DESCRIPTION
71129642
Indicator Bracket Assembly
71125105
Stainless Steel Column, 350 mm high
71125103
Stainless Steel Column, 680 mm high
71129064
Column Bracket
71123975
Load Cell, MT1241-50kg CH15R
71123976
Load Cell, MT1241-100kg CH30R, CH60R, CH60L
71123979
Load Cell, MT1241-250kg CH100R
71124310
Load Cell, MT1260-500kg CH150R, CH300R
71102504
Load Cell Mounting Bolts, M6 x 25
71102512
Load Cell Mounting Bolts, M8 x 230
71129104
Stainless Steel Platform, 305 x 355
71129105
Stainless Steel Platform, 400 x 500
71129106
Stainless Steel Platform, 420 x 550
71125079
Platter Rubber Supporter
71125924
Corner Rubber Supporter, 305 x 355
71125006
Corner Rubber Supporter, 400 x 500 & 420 x 550
71127181
Cover for M6 Screw
71128342
Cover for M8 Screw
71125009
Adjustable Foot (All)
Summary of Contents for CHAMP II BASE