EN-22
Adventurer Pro
Defining Custom Unit
Set Custom to ON in the Unit menu to enable
and define the Custom Unit. The custom unit
is defined by a conversion factor and a least
Significant Digit (LSD). The conversion factor
is used by the balance to convert grams to the
custom weighing unit and is defined by entering
a factor and an exponent. The factor is a value
between 0.1000000 and 1.999999 inclusive.
The exponent moves the decimal point of the
factor to the right for positive values or to the
left for negative values.
Enter the Factor, following the instructions for
Editing Limits in Section 3.5.4. Enter the
Exponent and LSD using the
Yes
No
buttons.
Factor
Exponent
Conversion
(+3 to -3)
Factor
.1234
3
123.4
.1234
2
12.34
.1234
1
1.234
.1234
0
.1234
.1234
-1
.01234
.1234
-2
.001234
.1234
-3
.0001234
3.7.5 Unit (Cont.)
Custom Unit = Conversion Factor x Grams.
The LSD is the value by which the displayed
weight is incremented or decremented.
LSD
Result
.5
Adds one decimal place
Display counts by 5
1
Display counts by 1
2
Display counts by 2
5
Display counts by 5
10
Display counts by 10
100
Display counts by 100
End Unit
Press the
Yes
button to advance to the Print-1
submenu. Press
No
to return to the first
available unit.
PRINT-1
Output When Stable
On/
Off
GLP Tare
On/
Off
Auto Print
Off
, Cont., Interval, When Stable
1 -3600 seconds
Stable- Load, Load & Zero
Content Num Only-
Off
, all others
On-
Header,Gross,Net, Tare,
Ref, Result,GLP
Layout Line Format -
Multi
, Single
4LF-Yes/
No
Form Feed-Yes/
No
List
Yes/
No
End Print-1
3.7.6 Print-1 and Print-2
The Print-1 submenu is used to set printing
parameters for an external printer or computer.
Print-2 contains an identical menu and is used
when the optional COM2 port is installed.
Output
Set When Stable to On to print only stable
values. Set When Stable to Off to print stable
or unstable values.
Set GLP Tare to On to print GLP data once after
a tare operation. Set GLP Tare to Off to disable
this feature.
Auto Print
When set to Continuous, the displayed value
is printed continuously. When set to Interval,
the displayed value is printed at the user
specified time interval (1 to 3600 seconds).
If set to When Stable, the balance will
automatically print the displayed value when
stability is achieved. An additional setting must
be made to determine if only stable non-zero
values will be printed (load setting) or if stable
zero and non-zero values will be printed (Load
& Zero). When set to Off, the Auto print feature
is disabled.
Summary of Contents for Adventurer Pro AV811 Series
Page 1: ...Adventurer Pro Balances Instruction Manual Adventurer Pro...
Page 41: ......
Page 45: ...5 1 1 51g 8100g 1 2 l AC 2 l l l RS232 l l 1 3 l AC AC l l l l l...
Page 46: ...6 2 2 1 51g 51g 210g AC 2 2 2 3 2 3 4...
Page 48: ...8 3 3 1 No No 1 8 2 9 AC 3 10 4 11 5 12 6 13 7 14...
Page 52: ...12 3 4 3 4 1 End Mode...
Page 56: ...16 3 5 3 Function 50g Mode PERCENT 3 5 4 3 4 3 EDITREF No EDIT REF Yes UNDER OVER Yes No...
Page 59: ...19 3 5 6 Function Mode TOTALIZE 3 6 3 6 1 NOTE...
Page 61: ...21 3 7 3 4 2 3 7 1 2 2 25 100 3 Note...
Page 69: ...29 3 10 RS232 1...
Page 73: ...33 5 1...
Page 76: ...36 5 3 2 RS232 9 D COM PRINT TARE...
Page 77: ...37 FAX BW...
Page 78: ...FR 38 Adventurer Pro...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......