ADVENTURER BALANCES
EN-41
7.
PRINTING
7.1
Connecting, Configuring and Testing the Printer/Computer Interface
Use the built-in RS-232 Port to connect either to a
computer or a printer.
If connecting to a computer, use HyperTerminal or similar software like SPDC described
below.
(Find HyperTerminal under
Accessories/Communications
in Windows XP.)
Connect to the computer with a standard (straight-through) serial cable.
Choose
New Connection
, “connect using” COM1 (or available COM port).
Select
Baud=9600
;
Parity=8 None
;
Stop=1
;
Handshaking=None
. Click
OK
.
Choose Properties/Settings, then ASCII Setup. Check boxes as illustrated:
(
S
end line ends…
;
E
cho typed characters…
;
W
rap lines…
)
Use RS232 Interface Commands (Section 9.6.1) to control the balance from a PC.
Note:
When the HyperTerminal configuration is complete, it will automatically
print the results of a
Cal Test
operation, and echo print commands sent to the scale.
SPDC Software
The Serial Port Data Collection or SPDC software is provided by Ohaus and can be used on
operating systems that do not have the HyperTerminal software mentioned above.
Choose the export file type and export file path and then press Run as shown below.
Note:
The SPDC software only supports English language.
7.2
Output Format
The Result Data, and G/N/T data, is output in the following format.
Field:
Label
1
Space
2
Weight
3
Space
2
Unit
4
Space Stability
5
Space G/N
6
Space
Term.
Characters
7
Length:
1
11
1
5
1
1
1
3
0
8
1. The length of the label field is not fixed.
2. Each field is followed by a single delimiting space (ASCII 32).
3. The Weight field is 11
right justified characters. If the value is negative, the “–“ character is located at
the immediate left of the most significant digit.
4. The Unit field contains the unit of measure abbreviation up to 5 characters, right justified.
5. The Stability field contains
the “?” character if the weight reading is not stable. The Stability field and the
following Space field are omitted if the weight reading is stable.
6. The G/N fi
eld contains the net or gross indication. For net weights, the field contains “NET”. For gross
weights,
the field contains nothing, “G”.
7. The Termination Characters field contains CRLF, Four CRLF or Form Feed (ASCII 12), depending on
the LINE FEED menu setting.
Summary of Contents for Adventurer AX124
Page 1: ...AdventurerTM Balances Instruction Manual ...
Page 2: ......
Page 6: ...EN 4 ADVENTURER BALANCES ...
Page 64: ......
Page 65: ...Balanzas AdventurerTM Manual de instrucciones ...
Page 66: ......
Page 128: ......
Page 129: ...BalancesAdventurerTM Manuel d instructions ...
Page 130: ......
Page 192: ......
Page 193: ...AdventurerTM Waagen Bedienungsanleitung ...
Page 194: ......
Page 257: ......