background image

UNPACKING AND SET-UP (Cont.) 

7.  Place the Scale Plate on the four vertical Support Posts, with the solid weld 

stud resting on the left front Post (with the two Lock Nuts). Hook the mass 
Hanger Assembly on the Ring that is located at the right end of the Beam. 

8.  Unpack the Slotted Masses from the corrugated box and place them on the  

Mass Rack. 

USING THE BALANCE (ZEROING) 

1.  Move all four Poises to their ZERO positions at 

the extreme left end of travel.  The Rear Poise is 
equipped with a pin that will position it 
accurately in the notches.   

2.  If the Balance does not come to rest exactly at 

zero, it can be rebalanced by using the two 
knurled, brass Adjustment Nuts of the Balance 
Compensator Assembly. When correct balance 
has been restored, be sure to lock the Balance 
Nuts against each other to prevent movement.

 

 

Balance Compressor 

WEIGHING 

Net Weight can be determined rapidly by following
these steps: 

1.  Counterbalance the EMPTY container with the 

two Tare Poises. Loosen the Thumb Screw of 
the larger Tare Poise and move both Poises to 
the right.  The larger Tare Poise is used for 

COARSE

 balancing and then is locked with the 

thumb screw.  The smaller Tare Poise is used 
for 

FINE

 balancing.  It is spring-loaded to protect

against accidental movement. 

 

  

Tare Poise 

 

2. Place the specimen as near the center of the Scale Plate as possible. 
3. Move the Rear Poise to the right until it is in the first notch that will cause the 

Beam to DROP below the Balance Indicator at the pointer end of the beam.  
Move it to the left one notch and the Beam will RISE. 

4. Move the Front Poise to the right until balance is obtained. 
5. Read the weight of the specimen directly by adding the values indicated by 

the Poises. 

6. If the Rear Poise can be moved to the extreme right notch but it does not 

cause the Beam to DROP, return both the Front and Rear Poises to ZERO 
and add Slotted Attachment Masses to the Mass Hanger Assembly.  Add 
sufficient masses to cause the Beam to DROP below the Balance Indicator 
at the pointer end of the Beam.  Remove the smallest (lightest) mass and 
follow Steps 3, 4 and 5.  Add the total value of the Attachment Masses on 
the Mass Hanger Assembly to the values indicated by the positions of the 
two poises to obtain weight of the specimen. 

Reviews: