Malfunctions and possible causes 

 

Problem 

To do 

1.

 

Signaling that the threshold value has been exceeded with normal soft water hardness 
below the limit value

 

to low water flow through measurement 
chamber 

 

when the valve is switched on, water must 
come from the drain 

the raw water valve does not close com-
pletely (e.g. due to deposits in the valve 
seat), which means that small amounts of 
raw water flow into the measuring cell in 
addition to the soft water 

replace valve 

low brine concentration in soft water

 

rinse active column of softening plant once 
again

 

sensor possibly contaminated  
or damaged by brine 

replace sensor 

2.

 

Sensor not recognized 

sensor not plugged in correctly  
or electrical plug connection defective 

Check plug connection or connection cable 
for continuity 

3.

 

Brine alarm 

high concentration of brine in the soft wa-
ter, conductivity compared to the raw wa-
ter is significantly higher

 

rinse the active column again 

 

raw or soft water flow interrupted, sensor 
no longer rinsed with soft or raw water 

restore the water supply 

sensor possibly contaminated or damaged 
by brine  

replace sensor 

4.

 

Valve leaking 

 

one of the two valves no longer closes 
properly

 

remove the hoses from the measurement 
chamber and check whether drops of water 
are escaping despite the closed valves  

 replace affected valve

 

5.

 

Raw or soft water does not flow 

sample inlet closed 

 

check shut-off valves 

 

pre-filter blocked 

clean fine-filter 

sample hose blocked 

check sample hose for flow 

defective valve 

replace valve 

Summary of Contents for Softcontrol Smart

Page 1: ...Softcontrol Smart Instruction Manual Version 2 3...

Page 2: ......

Page 3: ...SIGNALING VIA POTENTIAL FREE CONTACT 13 9 3 2 CONNECTION EXTERNAL WATER METER 13 9 3 2 1 CONNECTION MECHANICAL WATER METER 13 9 3 2 2 CONNECTION ELECTRONICAL WATER METER TURBINE WITH HALL CONTACTS 13...

Page 4: ...4...

Page 5: ...mponents improper analytical mechanical or electrical impact of any kind damage or consequential damage caused by improper use or operation damage or consequential damage caused by faulty or improper...

Page 6: ...ortance for a safe and hazard free operation of the device Attention Failure to comply with instructions marked in this way may result in damage to the device or the sensors Warning If notices marked...

Page 7: ...w voltage The raw and soft water of the softening process to be monitored must basically meet the requirements of the applicable drinking water ordinance A constant sample pressure of a minimum of 1 t...

Page 8: ...ly every 15 minutes in a flow measuring chamber in which a selective sensor for calcium and magnesium ions is inserted pres sure proof Alternatively the measuring cycle can be controlled by pulses fro...

Page 9: ...ration Hard water alarm brine intrusion If the water hardness is detected in the soft water the visual alarm is triggered immediately and the external hard water alarm is triggered after a delay The d...

Page 10: ...cated by an arrow on the filter housing The fine filters are preferably installed directly at the sam pling points for raw and soft water in order to also protect the sample hoses from clogging 2 Inse...

Page 11: ...raw water optional connection water meter valve raw water measurement chamber power supply 240V 50 Hz V softener filter filter raw water sensor cable drain pressure free 1 1 2 3 5 6 4 sensor soft wate...

Page 12: ...installation Optionally a water meter or an external signal output can be connected using the supplied 5 pin plug 1 2 3 4 N ot e E ac h te r m in al is m ar k e d w it h a n 5 S of t w at er Note The...

Page 13: ...d to hide excess hardness lev els due to so called hardness it is recommended not to carry out the monitoring according to fixed time cycles but continuously during the entire soft water withdrawal To...

Page 14: ...e time LED OK measurement lights up passive operation mode no increased water hardness recog nized good soft water quality LED OK measurement is flashing aktive operational mode measurement or calibra...

Page 15: ...mly seated on the sensor and steamed up inside A comparison of the number on the sensor with the number given on the packaging and on the delivery note is recommended The sensor should be stored in th...

Page 16: ...ng chamber in which the sensor is located and at the same time pull the used sensor straight out of the flow measuring chamber when changing sensors 4 Take the new sensor out of the folding box and pu...

Page 17: ...nsor depending on its intensity and duration Strong deep black discolorations on the metal elec trodes indicate a high level of chlorine exposure The sensor can indicate that the limit value has been...

Page 18: ...ensor not plugged in correctly or electrical plug connection defective Check plug connection or connection cable for continuity 3 Brine alarm high concentration of brine in the soft wa ter conductivit...

Page 19: ...nsor protection cap filled with approx 1 ml of raw water carefully placed on the hardness sensor The sensor should be stored in a cool 10 15 C and dark place 14 Dismantling For dismantling the measuri...

Page 20: ...or connection of water meter and external signal output 4 m flexible pressure proof PA line 4 mm OD 2 mm ID sampling hose 1 m flexible pressure proof PA line 8 mm OD 6 mm ID drainage hose 2 pieces scr...

Page 21: ...x 4 watts Raw Soft water connection branch with connection option for PA hose OD 4 mm ID 2 mm particle condition 0 1 mm Raw Soft water pressure minimum 1 bar maximum 7 bars pressure difference as low...

Page 22: ......

Page 23: ...ezogen For verification in accordance with Low Voltage Directive 2014 35 EU the following standards were applied EN 61010 1 2010 Zum Nachweis entsprechend EMV Richtlinie 2014 30 EG wurden folgende Nor...

Page 24: ...OFS Online Fluid Sensoric GmbH Heidelbergweg 9 07580 Ronneburg Germany Manual Softcontrol Smart Version 2 3 Issue 11 2020 Subject to technical changes...

Reviews: